|
Model |
AN9651B(F) (6 in 1) |
|
Display Mode |
240℅128 LCD Large Screen |
|
Isolated Transformer Capacity |
Single phase:2000W~6000W |
|
Over-current Protection Limit |
When current 30A,exceed 5s protection |
|
Alarm Mode |
Alarm with sound and light |
|
Standard Interfaces |
RS232 interface, Alarm Light interface, Remote Control interface,AN97 Frequency Converter Power Supply interface |
|
Interfaces Optional |
RS485 interface, Printer interface, Foot Pedal interface, On-off interface |
|
Buzzer Alarm Sound |
Can ON and OFF |
|
Test Group |
15 Group,7 test steps/group |
|
Ground Bond Test |
No-load Output Voltage |
<12V |
|
Current Presetting Range |
AC 3A~30A 50Hz/60Hz 1A/step |
|
Current Accuracy |
㊣(1.5% of setting+0.40A) |
|
Current Stability |
<0.4% (10A,100m次) |
|
Resistance Range/Accuracy |
10~600m次 ㊣(1.5% of reading+1.5% of Range) |
|
Resistance Upp-Limit Range |
3A≒Current≒10A:10~600m次; 11A≒Current≒25A:10~300m次;
26A≒Current≒30A:10~200m次 1m次/step |
|
Resistance Low-Limit Range |
0m次 |
|
Compensation Resistance Range |
0~200m次,Automatic Measurement, ON/OFF |
|
Voltage Range/Accuracy |
30~7500mV ㊣(1.5% of reading+1.5% of Range) |
|
Voltage Upp-Limit Range |
30~7500mV 1mV/step |
|
Voltage Low-Limit Range |
0mV |
|
Compensation Voltage Range |
0~1000mV,Automatic Measurement, ON/OFF |
|
Test Time Range/Resolution |
1~999s 1s/step |
|
Insulation Resistance Test |
Rated Output Load/Current |
2VA(1000V/2mA) 2mA |
|
Output Voltage Range/Resolution |
DC 500V/1000V <1% (From no load to full load) |
|
Output Voltage Ripple |
<1% (Pure Resistive 1000V/1mA) |
|
Output Voltage Accuracy/Resolution |
㊣(3% of setting+3V) 5V |
|
Resistance Range/Accuracy |
≒200M次: ㊣(3% of reading+3 counts);
>200M次: ㊣(8% of reading+8 counts) |
|
Resistance Upp-Limit Range |
2000M次 |
|
Resistance Low-Limit Range |
0.3~2000M次 0.1M次/step(≒100M次),1M次/step(>100M次) |
|
Test Time Range/Accuracy |
1~999s 1s/step |
|
AC Hipot Test |
Rated Output Load |
500VA(AC:5000V/100mA) |
|
Output Voltage Range/Freq/Resolution |
500~5000V 50Hz/60Hz <3%(From no load to full load) |
|
Output Voltage Setting Range/Accuracy |
Static:AC,500~5000V;Dynamic:AC,500~2500V;
㊣(3% of setting+10V) |
|
Breakdown Current Meas. Accuracy |
㊣(3% of reading+0.03mA) |
|
Breakdown Current Alarm Preset Range |
Upp-limit:0.10-100.00mA;Low-limit:0.03-10.00mA; 0.01mA/step |
|
Compensation Current Range |
0.00-10.00mA,Automatic Measurement, ON/OFF |
|
Test Time Range |
1~999s 1s/step |
|
Rise Time Range |
0~100s 1s/step |
|
;Leakage Current Test |
Output Voltage Range |
1.06 times of input voltage(output by isolated transformer) |
|
Leakage Mode Setting |
Static, Dynamic |
|
Output Voltage Meas. Range/Accuracy |
AC 60~280V/45~65Hz; ㊣(0.5% of reading+2 counts) |
|
Leakage Current Setting Range |
Upp-limit:0~9.999mA 0.001mA;Low-limit:0mA |
|
Compensation Current Setting Range |
0~1000uA,Automatic Measurement, ON/OFF |
|
Leakage Current Meas. Range/Accuracy |
0.050mA~9.999mA/
DC~10kHz: ㊣(3% of reading+10uA);
10k~1MHz: ㊣(5% of reading+50uA) |
|
Test Time Range |
1~999s 1s/step |
|
Power Test |
Output Voltage Range |
1.00 times of input voltage(output by isolated transformer) |
|
Power Setting Range/Resolution |
Upp-limit:6~6000W Low-limit:0~6000W 1W/step |
|
Output Voltage Meas. Range/Accuracy |
AC 60~280V; 45~65Hz: ㊣(0.5% of reading+2 counts) |
|
Current Meas. Range/Accuracy |
AC 0.030~3.999A,4.00~25.00A;
45~65Hz: ㊣(0.5% of reading+0.020A) |
|
Real-Power Meas. Range/Accuracy |
10.0~199.9W: ㊣(0.5% of reading+5W)
200~6000W: ㊣(0.5% of reading+30W) |
|
Power Factor Meas. Range/Accuracy |
0.100~1.000; ㊣(2% of reading+2 counts) |
|
Test Time Range |
1~999s,1s/step |
|
Low Voltage Start Test |
Output Voltage Range |
1.00 times of input voltage(output by isolated transformer) |
|
Current Meas. Range/Resolution |
Upp-limit:0.30~25.00A Low-limit:0.00~25.00A 0.1A |
|
Output Voltage Meas. Range/Accuracy |
AC 60~280V; 45~65Hz: ㊣(0.5% of reading+2 counts) |
|
Current Meas. Range/Accuracy |
AC 0.030~3.999A,4.00~25.00A
45~65Hz: ㊣(0.5% of reading+0.010A) |
|
Test Time Range |
1~999s 1s/step |