HOME>> Product >> Electrical Safety Comprehensive System >> Electrical Safety Comprehensive Tester :
Product

Electrical Safety Comprehensive Tester:AN9641B(F)

Sum Mary :

Download :

Product Infoㄩ

Main Features:

↑Compliant with IEC60335-1,GB4706.1-2005,GB4943 standard requirements;
↑Measurement Accuracy of electrical safety test is 3% and Measurement Accuracy of power test is 0.5%;
↑Can preset 15 groups,7 items per group;
↑Delete former forcibly self-check function,add compensation function to compensate GB,IR,ACW,LC test at any time,more convenient for users;
↑Standard device for appliance test,suitable for testing more than 800 appliance production lines;
↑Blue LCD display,all measurement parameters are displayable;
↑Enhanced communication with remote control and test results can be saved instantly;
↑Steady output supply voltage and current for all tests;
↑Supply 50Hz and 60Hz frequency,a highly integrated electrical safety test instrument
↑Enhanced test analysis,memory,alarm setting,error offset,and auto protection functions
↑Built-in isolated transformer to improve the reliability & safety of operation;
↑Equip with custom design high speed test programs

Technical Parameter:

Model
AN9641B(F) (4 in 1)
Display Mode
240℅128 LCD Large Screen
Isolated Transformer Capacity
Single phase:2000W~6000W
Over-current Protection Limit
When current 30A,exceed 5s protection
Alarm Mode
Alarm with sound and light
Standard Interfaces
RS232 interface,Alarm Light interface,Remote Control interface,AN97 Frequency Converter Power Supply interface
Interfaces Optional
RS485 interface,Printer interface,Foot Pedal interface,On-off interface
Buzzer Alarm Sound
Can ON and OFF
Test Group
15 Group,7 test steps/group
              Ground Bond Test
No-load Output Voltage
<12V
Current Presetting Range
AC 3A~30A 50Hz/60Hz 1A/step
Current Accuracy
㊣(1.5% of setting+0.40A)
Current Stability
<0.4% (10A,100m次)
Resistance Range/Accuracy
10~600m次 ㊣(1.5% of reading+1.5% of Range)
Resistance Upp-Limit Range
3A≒Current≒10A:10~600m次; 11A≒Current≒25A:10~300m次;
26A≒Current≒30A:10~200m次  1m次/step
Resistance Low-Limit Range
0m次
CompensationResistance Range
0~200m次,Automatic Measuremet,ON/OFF
Voltage Range/Accuracy
30~7500mV ㊣(1.5% of reading+1.5% of Range)
Voltage Upp-Limit Range
30~7500mV 1mV/step
Voltage Low-Limit Range
0mV
Compensation Voltage Range
0~1000mV,Automatic Measurement,ON/OFF
Test Time Range/Resolution
1~999s 1s/step
Insulation Resistance Test
Rated Output Load/Current
2VA(1000V/2mA) 2mA
Output Voltage Range/Resolution
DC 500V/1000V <1% (From no load to full load)
Output Voltage Ripple
<1% (Pure Resistive 1000V/1mA)
OutputVoltage Accuracy/Resoluton
㊣(3% of setting+3V) 5V
Resistance Range/Accuracy
≒200M次: ㊣(3% of reading+3 counts);
>200M次: ㊣(8% of reading+8 counts)
Resistance Upp-Limit Range
2000M次
Resistance Low-Limit Range
0.3~2000M次 0.1M次/step(≒100M次),1M次/step(>100M次)
Test Time Range/Accuracy
1~999s 1s/step
      AC Hipot Test
Rated Output Load
500VA(AC:5000V/100mA)
OutputVoltageRange/Freq/Resolution
500~5000V 50Hz/60Hz <3%(From no load to full load)
OutputVoltageSettingRange/Accuracy
Static:AC,500~5000V;Dynamic:AC,500~2500V;
㊣(3% of setting+10V)
Breakdown Current Meas.Accuracy
㊣(3% of reading+0.03mA)
BreakdownCurrentAlarmPresetRange
Upp-limit:0.10-100.00mA;Low-limit:0.03-10.00mA; 0.01mA/step
Compensation Current Range
0.00-10.00mA,Automatic Measurement,ON/OFF
Test Time Range
1~999s 1s/step
Rise Time Range
0~100s 1s/step
Leakage Current Test
Output Voltage Range
1.06 times of input voltage(output by isolated transformer)
Leakage Mode Setting
Static,Dynamic
OutputVoltage Meas.Range/Accuracy
AC 60~280V/45~65Hz; ㊣(0.5% of reading+2 counts)
Leakage Current Setting Range
Upp-limit:0~9.999mA 0.001mA;Low-limit:0mA
CompensationCurrentSettingRange
0~1000uA,Automatic Measurement,ON/OFF
LeakageCurrentMeas.Range/Accuracy
0.050mA~9.999mA/
DC~10kHz: ㊣(3% of reading+10uA);
10k~1MHz: ㊣(5% of reading+50uA)
Test Time Range
1~999s 1s/step

 

Previousㄩ No more products
NextLeakage Current Tester:AN9620TH(F)