Model |
AN9634H(F)
AC/DC Hipot |
|
AC Hipot Test |
Output Voltage Setting / Test Range |
100V~5000V AC 100mA 1V/step |
|
Output Voltage Adjustments |
¡À(2% of setting+5V) |
|
Output Voltage Resolution/ Accuracy |
10V, ¡À(2% of setting+5V) |
|
Output Frequency Range/ Accuracy |
50Hz/60Hz, ¡À100ppM |
|
Output Voltage Waveform |
Sine wave for pure resistive load , > 200V, distortion <2%, 100V-200V< 5% |
|
Breakdown Current Alarm Upper and Lower Limit Setting Range |
Upper limit 0.00~99.99mA, 0.01mA/step
Lower limit 0.000~9.999mA, 0.001mA/step |
|
Breakdown Current Range |
0.000~3.500mA/3.50~99.99mA |
|
Breakdown Current Resolution |
0.001mA/0.01mA |
|
Breakdown Current Accuracy |
< ¡À(2% of reading+3counts) |
|
Rise Time /Fall time/Test Time Range |
0.0~999.9s/0.0~999.9s 0.1s/step |
|
Offset Cancellation Function |
0~2.000mA, Auto or manual offset compensation |
|
DC Hipot Test |
Output Voltage Setting and Range |
100V~6000V DC 10mA 1V/step |
|
Output Voltage Regulated Rate |
¡À(2% of setting+5V) |
|
Resolution and Accuracy |
10V, ¡À(2% of setting+5V) |
|
DC Ripple Quotient |
No load¡Ü2%, Full load¡Ü5%(under 6kV DC/10mA) |
|
Breakdown Current Alarm Upper and Lower Limit Setting Range |
Upper limit 0.0~9999¦ÌA, 1¦ÌA/step
Lower limit 0.000~999.9¦ÌA , 0.1¦ÌA/step |
|
Upper//Low Limit Setting Range of DC Rising Current |
12mA, ON / OFF optional / 0.0~350.0¦ÌA |
|
Breakdown Current Measurement Range |
0.0~350.0¦ÌA /300~3500¦ÌA/3.00~9.99mA |
|
Breakdown Current Measurement Resolution & Accuracy |
0.1¦ÌA /1¦ÌA /0.01mA <¡À(2% of reading+3counts) |
|
Rise-Time Setting and Measurement Time Range |
0.0~999.9s/0.5~999.9s 0.1s/step |
|
Offset Cancellation Function |
0~200¦ÌA, Auto or manual offset compensation |
|
Dimension(W¡ÁH¡ÁD) (mm) |
Desktop style:400¡Á156¡Á475 |