HOME>> Product >> Electrical Safety Tester >> Hipot / Insulation Resistance Tester :
Product

AC Hipot/Insulation Resistance Tester:AN9632X(F)

Sum Mary :

Download :

Product Info£º

Main features
¡ôVoltage ramp-up/ramp-down function, accords with new standard requirements
¡ôUpper and lower limit leakage current preset function, intelligent judgment, voice and light alarm
¡ôSupports both 50Hz and 60Hz
¡ôOver voltage, over current protection, Test line clears to zero by Manual
¡ôAdopt VFD display, simple operation
¡ôRemote interface, RS232/485 interface function
¡ô Smaller size and more beautiful appearance.


Technical parameters

Model
AN9632X
AC Hipot Test
Rated Output
100VA  (AC: 5kV/20mA)
Output  Voltage Set and Range
0.2kV¡«5.00kV AC,  10V/step
Output Voltage Adjustable Rate
¡À(2.5% of setting+10V)
Output Voltage Resolution/Accuracy
10V, ¡À(2.5% of setting+10V)
Output Frequency Range
50Hz, 60Hz (Optional)
Output Waveform
Sine waveform, pure resistance, above 200V, distortion <5%
Upper and Lower Limit Leakage Current Setting
Upper: 0.10¡«20.00mA,  Lower:0.10¡«20.00mA(0=disable), 0.01mA/step
Leakage Current Measurement Range
0.100¡«3.499mA/3.50¡«20.00mA
Leakage Current Resolution
0.001mA/0.01mA
Leakage Current Accuracy
¡À(2.5% of reading+5 counts)
Ramp-up Time Setting Range/Resolution
1.0¡«999.9s (0=disable)   0.1s/step
Test Time Setting Range/Resolution
1.0¡«999.9s (LLL=continues)   0.1s/step
DUT Abnormal Processing
Buzzer alarm, Light alarm
Test Line Clear to Zero
0¡«0.500mA, manual clear offset
Insulation Resistance Test
Output Voltage Setting and Measurement Range
100V ¡«1000V  DC  5V/step 
Output Voltage Resolution/Accuracy
1V , ¡À(3% of setting+5V)
DC Ripple Coefficient
¡Ü5% (1kV DC, <1mA, resistance load)
Output Voltage Adjustable Rate
¡À(3% of setting+5V)
Test Time Setting Range
1.0¡«999.9s 0.1s/step (LLL= continues)
Lower limit of Insulation Resistance
1¡«2000M
Upper limit of Insulation Resistance
1¡«2000M (0=off)
Insulation Resistance Range
1¡«2000M, auto shift level
Insulation Resistance Accuracy
500¡«1000V: ¡À(3% of reading+8counts) 1¡«200M;
¡À(8% of reading+8counts) 200¡«2000M
Below 500V: ¡À(5% of reading+8counts) 1¡«200M;
¡À(10% of reading+8counts) 200¡«2000M
General Parameter
Interface Function
RS232(RS485 optional), Remote control(PLC)
Dimension  W¡ÁH¡ÁD(mm)
320¡Á110¡Á400

Previous£º AC Hipot/Insulation Resistance Tester:AN9632H(F)