Model |
AN9632H(F)
AC Hipot/Insulation Resistance |
|
AC Hipot Test |
Output Voltage Setting / Test Range |
100V~5000V AC 100mA 1V/step |
|
Output Voltage Adjustments |
¡À(2% of setting+5V) |
|
Output Voltage Resolution/ Accuracy |
10V, ¡À(2% of setting+5V) |
|
Output Frequency Range/ Accuracy |
50Hz/60Hz, ¡À100ppM |
|
Output Voltage Waveform |
Sine wave for pure resistive load , > 200V, distortion <2%, 100V-200V< 5% |
|
Breakdown Current Alarm Upper and Lower Limit Setting Range |
Upper limit 0.00~99.99mA, 0.01mA/step
Lower limit 0.000~9.999mA, 0.001mA/step |
|
Breakdown Current Range |
0.000~3.500mA/3.50~99.99mA |
|
Breakdown Current Resolution |
0.001mA/0.01mA |
|
Breakdown Current Accuracy |
< ¡À(2% of reading+3counts) |
|
Rise Time /Fall time/Test Time Range |
0.0~999.9s/0.0~999.9s 0.1s/step |
|
Offset Cancellation Function |
0~2.000mA, Auto or manual offset compensation |
|
Insulation Resistance Test |
Output Voltage Setting and Range |
100V~1000V DC 1V/step |
|
Output Voltage Resolution andAccuracy |
1V, ¡À(2% of setting+5V) |
|
DC Ripple Quotient |
No load¡Ü1% ,Full load<5%(under 1K DC/1mA) |
|
Output Voltage Regulated Rate |
¡À(2% of setting+5V) |
|
Test time Setting Range |
0.5~999.9s 0.1s/step |
|
Lower limit Setting Range of Insulation Resistance |
1~9999M |
|
Upper limit Setting Range of Insulation Resistance |
0~9999M |
|
Lower Limit Setting Range of DC Charging Current |
0.000~3.500¦ÌA |
|
Insulation Resistance Range |
1~9999M, automatic gear shift,4 digit display |
|
Insulation Resistance Measurement Accuracy |
500~1000V: ¡À(2% of reading+2counts)1~200M;
¡À(5% of reading+2counts)200~1000M,
¡À(15% of reading+2counts)1000~9999M,
Below 500V: ¡À(8% of reading +2counts) 1~1000M
¡À(15% of reading +2counts) 1000~2000M |
|
Dimension(W¡ÁH¡ÁD) (mm) |
Desktop style:400¡Á156¡Á475 |