Product Info£º
¡ôOriginal plug and play MD clip, display MD network circuit;
¡ôOriginal MD network circuit video, changeable, can be measured;
¡ôInner 4 kinds of main MD slip, can equip other slips;
¡ôDC-1MHz bandwidth, Earth Leakage Current, Enclosure leakage current and applied part leakage
¡ôCan test single phase/three phase test
¡ôIntelligent judgment, voice and light alarm, large font display;
¡ô8-group and 8-step test condition memory
|
MD |
Standard |
Application Area |
|
A |
L3101-1,UL1950,UL60950,IEC60950,IEC60065,IEC61010-1,IEC60335-1, IEC60990,GB/T12113,GB4943,GB8898,GB4706,GB7000 |
Examination and Communication equipments(apperceive and responses current) |
|
B |
UL1563,IEC60335-1 |
Electrical equipment and assembly products |
|
C |
UL2601,EN60601-1,IEC60601-1,GB9706.1 |
Medical instrument and equipment |
|
D |
UL544 Patient Care |
Medical instrument and equipment |
|
E |
UL544 Non Patient, UL484, IEC60598 |
Medical instrument and equipment |
|
F |
L3101-1,UL1950,UL60950,IEC60950,IEC60065,IEC61010-1,IEC60335-1,IEC60990,GB/T12113,GB7000 |
Examination and Communication equipments(slip current) |
|
G |
IEC60950-1 |
Function Test |
|
Model |
AN9620H(F) |
|
Power supply |
220V ¡À10%, 50Hz |
|
Test input Voltage |
Phase voltage: 30V ~ 300V AC single phase |
|
Test input Current |
23A AC MAX |
|
Voltage Display Range/ Resolution/ Accuracy |
Range: 30.0 ~ 300.0V, Resolution :0.1V, Accuracy:¡À(1% of reading+0.5V) |
|
Over Current Protect |
25A |
|
Inner MD |
MD A,C, E, F (default), other MD is optional |
|
Leakage Current Measurement Range |
MD A¡¢C¡¢D¡¢E¡¢F: 10uA¡«6000uA
MD B: 10uA¡«18.00mA
MD G: 10uA¡«4500uA |
|
Leakage Current Measurement Resolution |
10.0¦ÌA~999.9¦ÌA Resolution: 0.1¦ÌA
1000¦ÌA~9999¦ÌA Resolution: 1¦ÌA
10.00mA~18.00mA Resolution: 10¦ÌA |
|
Accuracy
|
MD:A¡¢B¡¢C¡¢D¡¢E¡¢F |
10uA¡«50uA DC¡«1M ¡À£¨2%of reading +3uA£©
50uA¡«MDmax DC¡«10k ¡À£¨2%of reading+0.5uA£©;
10k¡«200k ¡À£¨5%of reading+0.5uA£©;
200k¡«1M ¡À£¨5%of reading+1.5uA£© |
|
MD:G |
10uA¡«50uA DC¡«1M ¡À£¨2%of reading+12uA£©
50uA¡«100uA DC¡«1M ¡À£¨2%of reading+4uA£©
100uA¡«MDmax DC¡«10k ¡À£¨2%of reading+0.5uA£©;
10k¡«200k ¡À£¨5%of reading+0.5uA£©;
200k¡«1M ¡À£¨5%of reading+1.5uA£© |
|
Leakage Current Alarm Preset |
Upper Limit preset: 0~MAX of measurement range(0=disable)
Lower Limit preset: 0~MAX of measurement range(0=disable) |
|
Measurement Time Setting Range/ Resolution / Accuracy |
0~999.9s( 0=continues test); Resolution :0.1s Accuracy: ¡À(0.1% of setting +0.1s) |
|
Display Mode |
LCD |
|
Interface |
Remote control(PLC), RS232(RS485 optional) |
|
Dimension (W¡ÁH¡ÁD) (mm) |
400¡Á156¡Á475 |
AN9620H Manuals.pdf
|