HOME>> Product >> Electronic Load >> AN23 Series DC Electronic Load :
Product

DC Electronic Load:AN23101H(F)/23102H(F)/23103H(F)

Sum Mary :

Download :

Product Infoㄩ

Main Features:
↑Flexible combination mode,load can be configured according to needs;
↑Large LCD display,each test parameter can be displayed,Operation is fast and simple;
↑Constant voltage,constant current,constant resistance,constant power,dynamic load,short current mode,ect.
↑Adopt high frequency DDS(direct digital synthesis),to control rise/ fall load current slope independently;
↑Adopt DSP (digital signal processing)technology and high precision 16-bit ADC,fast and accurate. Voltage tester,current tester,power tester,timer tester are all 5-digit display,can also test instantaneous short current;
↑Have HI and LO voltage,current,power GO/NG,and judgment function;
↑Multifold programmed load start/stop function;
↑Remote voltage test Vsense BNC input interface,can be used for exact voltage test,eliminate voltage test error of down-lead voltage drop when current is large;
↑Isolated current monitor BNC output interface;
↑Protection functions are well designed,contain over voltage,over current,over power,over temperature,and reverse polarity input protection;
↑Primary load has 150 groups of memory function,secondary loads has 50 groups memory function;
↑Multiple interface functions,easy to operate with remote control function;
↑Panel operation software precision emendation;
↑Fan rotation speed can be auto adjusted according to load power so that noise and power consumption are reduced

Model
AN23101H(F)
AN23102H(F)
AN23103H(F)
Input
 Range
Voltage
0~60V
Current
0~15A
0~30A
0~60A
Power
0~75W
0~150W
0~300W
Constant
Current
Range
0.004~1.5A
1.5A~15A
0.008~3A
3A~30A
0.015~6A
6A~60A
Resolution
0.0004A
0.004A
0.0008A
0.008A
0.0015A
0.015A
Accuracy
㊣(0.2% of setting +0.1% of range)
Constant
Resistance
Range
0.20次~4次
4次~16k次
0.10次~2次
2次~8k次
0.05次~1次
1次~4k次
Resolution
1.00m次
0.06ms
0.50m次
0.13ms
0.25m次
0.25ms
Accuracy
㊣(0.2% of setting +0.1% of range)
Constant
Voltage
Range
1~ 60V
Resolution
0.015V
Accuracy
㊣(0.1% of setting +0.05% of range)
Constant
Power
Range
0.2~75W
0.4~150W
0.8~300W
Resolution
0.02W
0.04W
0.08W
Accuracy
㊣(0.5% of setting +0.2% of range)
Dynamic Load
Load Current
1~62.5mA/ us
2~125mA/ us
4~250mA/ us
Rise up/Fall Down Rate
10mA~0.625A/us
20mA~1.25A/us
40mA~2.5A/us
High/Low Correct Phase Time
100us~9.999s
Randorm Waveform Period
10Hz~10kHz
Accuracy
㊣10%
Load off
≡100k次 (60V)
Display
Accuracy
DC Voltage
㊣(0.1% of reading +0.05% of range)
Current
㊣(0.2% of reading +0.1% of range)
Power
㊣(0.5% of reading +0.2% of range)
Time
㊣(0.1% of reading +1 count)
Protection
Function
Over Voltage
105% Max input voltage
Over Current
105% Max input voltage
Over Power
105% Max input voltage
Over Temperature
85⊥
Reverse Polarity
Have Reverse Polarity function
Vsense Interface
Vsense remote voltage test BNC input interface(lead voltage compensation)
Imonitor interface
Imonitor monitors BNC output interface
Dimension(W℅H℅D)
168℅178℅435
Weight
About 10kg

 

PreviousCharger/adapter Automatic Test System:AN8062